APPLICATION NOTE NO. 74
Calculating Calibration Coefficients for Turner
Cyclops-7 Fluorometer
Revised February
2005
![]()
Application note in pdf format
The Cyclops-7 provides a 0 to 5 volt output voltage proportional to fluorescence. The Cyclops-7 is factory-configured at Turner to detect one of the following fluorophores:
| Fluorophore | Chlorophyll a | Rhodamine WT tracer dye | Phycocyanin | Phycoerythrin |
| Designation on instrument | C | R | P | E |
The user can customize the 0-5V Cyclops-7 range to correspond to the expected data range by changing the interface cable, thus improving data resolution. The ranges are:
| Cable Gain | Chlorophyll a (µg/l) | Rhodamine WT tracer dye (ppb) | Phycocyanin or Phycoerythrin (cells/ml) |
| X100 | 0 – 5 | 0 – 10 | 0 – 20,000 |
| X10 | 0 – 50 | 0 – 100 | 0 – 200,000 |
| X1 | 0 – 500 | 0 – 1000 | 0 – 2,000,000 |
Example: You expect a range of 0 – 30 µg/l for chlorophyll a fluorescence. If you use an X1 cable, the maximum voltage will be 0.3 V (= 30 µg/l / [500 µg/l / 5 V] ). This limits the resolution and multiplies the noise level of the instrument. Changing the range to 0 - 50 µg/l (using an X10 cable) provides the best results.
When interfacing a Cyclops-7 to a Sea-Bird CTD, Sea-Bird defines the concentration as:
concentration = (scale factor * voltage) + offset
The factory range can be used to calculate the factory default scale factor and offset (see page 2 for field calibration):
scale factor = Range / 5 V
offset = - scale factor * measured voltage at 0 concentration (i.e., blank voltage)
To determine the blank voltage, place the fluorometer, in its field configuration (with or without a flow-through cap, as applicable) in de-ionized water and take a voltage reading.
Change the range by changing the interface cable, as needed, before you set up the .con file in Sea-Bird software.
Chlorophyll a – Equations shown are for units of µg/l; other units available
chlorophyll a (µg/l) = (scale factor * voltage) + offset
where Scale Factor is in µg/l-volt and Offset is in µg/lRhodamine ‑ Equations shown are for units of ppb; other units available
Rhodamine (ppb) = (scale factor * voltage) + offset
where Scale Factor is in ppb/volt and Offset is in ppbPhycocyanin or Phycoerythrin – Equations shown are for units of cells/ml.
Phycocyanin or Phycoerythrin = (scale factor * voltage) + offset
where Scale Factor is in cells/mL-volt and Offset is in cells/mL
Note concerning Phycocyanin or Phycoerythrin units: The current software version does not allow the user to select cells/mL as output units for the Turner SCUFA. Calculate and input the scale factor in cells/mL-volt and the offset in cells/mL. Select any units in the software, and be aware when you view and process the data that the units are actually cells/mL.
Note: The software dialog box also shows grayed out fields for MX, MY, and B; these calibration parameters are applicable only when a Turner OBS has been selected for another voltage channel (intended for use with a Turner SCUFA with an optional turbidity channel).
Example of Chlorophyll a Concentration Calculation:
If
fluorometer scale factor = 10.5 µg/l-volts,
offset = - 0.03 µg/l, and
measured voltage from fluorometer = 3.65 volts,
Calculated concentration (µg/l) = (scale factor * voltage) + offset = (10.5 * 3.65) - 0.03 = 38.29 µg/l
While the nominal scale factor and offset based on factory range and your measured blank voltage can be used to obtain approximate values, field calibration for chlorophyll a, Phycocyanin, and Phycoerythrin is highly recommended. The relationship between fluorescence and these parameters is highly variable, and is not easy to determine in the laboratory. Species distribution, ambient light level, and health of the stock are just some of the factors that affect the relationship.
To accurately measure concentrations, perform calibrations on seawater samples with concentrations of plankton populations that are similar to what is expected in situ. Determine concentrations independently, and use those concentrations, as well as readings from the fluorometer (with or without a flow through cap, as applicable), to determine the correct Scale Factor. It is only through the use of these calibrations that a meaningful and accurate measure of these parameters can be obtained. The scale factor is correct as long as the condition of the plankton population does not change; the condition does change with season and geographic location.
To accurately measure the offset, perform a calibration on a blank (de-ionized water) sample, with the fluorometer in its field configuration (with or without a flow-through cap, as applicable).
See Turner’s Cyclops-7 manual for calibration details.
![]()