Service
Sales
Ask Us
WET Labs
Satlantic

What is the fluorescent stick that came with my meter used for?

Checking Meter Output with the Fluorescent Stick
ECO meters ship with a stick of fluorescent material with which you can check your meter's output response.
    •    With the meter on and outputting data, hold the stick approximately 1/2-in (1 cm) from the lit detectors. Be careful to NOT touch the sensor face, as it scratches easily.
    •    Pivot the stick to parallel with the meter's optics; the output in the  meter's "signal" column should increase to near maximum output as given on the meter's characterization or calibration sheet.
    •    As you move the stick away, the output will decrease.