SeapHOx™ Ocean CT(D)-pH-DO Sensor

SeapHOx™ Ocean CT(D)-pH-DO Sensor

The Sea-Bird Scientific SeapHOx™ combines the SeaFET™ pH sensor with the SBE 37-SMP-ODO MicroCAT CTD+DO sensor.

The Sea-Bird Scientific SeapHOxTM combines the Satlantic SeaFETTM pH sensor with the Sea-Bird Electronics SBE 37-SMP-ODO MicroCAT CTD+DO sensor. The SeapHOxTM allows for the integrated data collection of pH with the critical oceanographic and biological measurement of temperature, salinity, and oxygen. The integrated package also allows the SeaFETTM to take advantage of the SBE 37's pumped flow path and anti-fouling technology, extending deployment durations in some cases.

Features

  • Moored pH, Conductivity, Temperature, Pressure (optional), and Optical Dissolved Oxygen, at user-programmable 5-minute to 24-hour intervals
  • Integral pump
  • RS-232 or USB interface
  • Internal memory and batteries (can be powered externally)*
  • Industry-leading conductivity cell biofouling protection
  • SeaFETCom© Windows software package (setup, data upload, and data processing)
  • Field-proven MicroCAT family, with more than 10,000 instruments deployed
  • Maximum depth 50 m

* The instrument MUST carry internal batteries; external power may extend the deployment duration depending on the sampling regime.

Components

  • pH sensor is Ion selective field effect transistor type
  • Unique internal-field conductivity cell permits use of expendable anti-foulant devices, for long-term bio-fouling protection
  • Aged and pressure-protected thermistor has a long history of exceptional accuracy and stability
  • Optional strain-gauge pressure sensor with temperature compensation is available in two ranges
  • Oxygen sensor is field-proven, individually calibrated SBE 63 Optical Dissolved Oxygen sensor
  • Pump runs for each sample, providing improved pH, conductivity, and oxygen response, bio-fouling protection, and correlation of CTD and oxygen measurements

Options

  • RS-232 or USB interface
  • Optional strain-gauge pressure sensor
  • Wire mounting clamp and guide or brackets for mounting to a flat surface

Measurement

Conductivity 0 to 7 S/m (0 to 70 mS/cm)
Temperature -5 to 45 °C
Optional Pressure 20 or 100 m measurement range
Dissolved Oxygen 120% of surface saturation in all natural waters (fresh and salt)
pH 6.5 - 9.0 pH

 

Initial Accuracy

Conductivity ± 0.0003 S/m (0.003 mS/cm)
Temperature ± 0.002 °C (-5 to to 35 °C); ± 0.01 °C (35 °C to 45 °C)
Optional Pressure ± 0.1% of full scale range
Dissolved Oxygen larger of ± 3 μmol/kg (0.07 ml/L, 0.1 mg/L) or ± 2%
pH 0.02 pH

 

Typical Stability

Conductivity 0.0003 S/m (0.003 mS/cm) per month
Temperature 0.0002 °C per month
Optional Pressure 0.05% of full scale range per year
Dissolved Oxygen sample-based drift < 1 μmol/kg/100,000 samples (20 °C)
pH 0.003 pH/month

 

Resolution

Conductivity 0.00001 S/m (0.0001 mS/cm)
Temperature 0.0001 °C
Optional Pressure 0.002% of full scale range
Dissolved Oxygen 0.2 μmol/kg
pH 0.004 pH

 

Miscellaneous

Memory Capacity 4 GB
System Depth Rating 50 m
System Dimensions 74.17 cm x 28.25 cm x 12.90 cm | 29.2" x 11.12" x 5.08"

 

The list below includes (as applicable) the current product brochure, manual, and quick guide; software manual(s); and application notes.

For older SeapHOx quickstart guides, click here.

SeatermV2© is a terminal program launcher for setup and data upload of Sea-Bird instruments developed or redesigned in 2006 and later. The common feature of this generation of instruments is the ability to output status responses in XML. SeatermV2 is part of our Seasoft V2 software suite.
Version 2.6.1 released June 1, 2016
SeatermV2.6.1-b12.exe for Windows XP/Vista/7


SBE Data Processing© consists of modular, menu-driven routines for converting, editing, processing, and plotting of oceanographic data acquired with Sea-Bird profiling CTDs, thermosalinographs, and the SBE 16 and 37 families of moored CTDs. SBE Data Processing is part of our Seasoft V2 software suite.
Version 7.26.2 released September 6, 2016


How often do I need to have my instrument and/or auxiliary sensors recalibrated? Can I recalibrate them myself?

General recommendations:

  • Profiling CTD — recalibrate once/year, but possibly less often if used only occasionally. We recommend that you return the CTD to Sea-Bird for recalibration. (In principle, it is possible for calibration to be performed elsewhere, if the calibration facility has the appropriate equipment andtraining. However, the necessary equipment is quite expensive to buy and maintain.) In between laboratory calibrations, take field salinity samples to document conductivity cell drift.
  • Thermosalinograph — recalibrate at least once/year, but possibly more often depending on the degree of bio-fouling in the water.
  • DO sensor —
    — SBE 43 — recalibrate once/year, but possibly less often if used only occasionally and stored correctly (see Application Note 64), and also depending on the amount of fouling and your ability to do some simple validations (see Application Note 64-2)
    — SBE 63 — recalibrate once/year, but possibly less often if used only occasionally and stored correctly and also depending on the amount of fouling and your ability to do some simple validations (see SBE 63 manual)
  • pH sensor —
    — SBE 18 pH sensor or SBE 27 pH/ORP sensor — recalibrate at the start of every cruise, and then at least once/month, depending on use and storage
    — Satlantic SeaFET pH sensor — recalibrate at least once/year. See FAQ tab on Satlantic's SeaFET page for details (How often does the SeaFET need to be calibrated?).
  • Transmissometer — usually do not require recalibration for several years. Recalibration at the manufacturer’s factory is the most practical method.

Profiling CTDs:

We often have requests from customers to have some way to know if the CTD is out of calibration. The general character of sensor drift in Sea-Bird conductivity, temperature, and pressure measurements is well known and predictable. However, it is very difficult to know precisely how far a CTD calibration has drifted over time unless you have access to a very sophisticated calibration lab. In our experience, an annual calibration schedule will usually maintain the CTD accuracy to within 0.01 psu in Salinity.

Conductivity drifts as a change in slope as a result of accumulated fouling that coats the inside of the conductivity cell, reducing the area of the cell and causing an under-reporting of conductivity. Fouling consists of both biological growth and accumulated oils and inorganic material (sediment). Approximately 95% of fouling occurs as the cell passes through oil and other contaminants floating on the sea surface. Most conductivity fouling is episodic, as opposed to gradual and steady drift. Most fouling events are small and mostly transitory, but they have a cumulative affect over time. A severe fouling event, such as deployment through an oil spill, could have a dramatic but only partially recoverable effect, causing an immediate jump shift toward lower salinity. As fouling becomes more severe, the fit becomes increasingly non-linear and offsets and slopes no longer produce adequate correction, and return to Sea-Bird for factory calibration is required. Frequently checking conductivity drift is likely to be the most productive data assurance measure you can take. Comparing conductivity from profile to profile (as a routine check) will allow you to detect sudden changes that may indicate a fouling event and the need for cleaning and/or re-calibration.

Temperature generally drifts slowly, at a steady rate and predictably as a simple offset at the rate of about 1-2 millidegrees per year. This is approximately equal to 1-2 parts per million in Salinity error (very small).

Pressure sensor drift is also an offset, and annual comparisons to an accurate barometer to determine offset will generally keep the sensor within specification for several years, particularly as the sensors age over time.

How accurate is salinity measured by my CTD? What factors impact accuracy?

One of the reasons that this is not a simple question is that there are several factors to take into consideration regarding the error margin for practical salinity measurements. Salinity itself is a derived measurement from temperature, conductivity, and pressure, so any errors in these sensors can propagate to salinity. For example, Oour initial accuracy specification for the SBE 3plus temperature sensor and SBE 4 conductivity sensor on an SBE 9plus CTD is approximately equivalent to an initial salinity accuracy of 0.003 PSU (note that conductivity units of mS/cm are roughly equivalent in terms of magnitude to PSU).

However, another issue to consider is that this accuracy is defined for a clean, well-mixed calibration bath. In the ocean, some of the biggest factors that impact salinity accuracy are 1) sensor drift from biofouling or surface oils for conductivity in particular and 2) dynamic errors that can occur on moving platforms, particularly when conditions are rapidly changing, which will be true for all sensors that measure salinity. Sea-Bird provides recommendations, design features such as a pumped flow path, and data processing routines to align and improve data for the salinity calculation to account for thermal transients and hysteresis, and to match sensor response times.  Depending on the environment and the steepness of the gradient, and after careful data processing, this may continue to have an impact on salinity on the order of 0.002 PSU or more, for example. For more details, see Application Note 82.

Lastly, note that salinity in PSU is calculated according to the Practical Salinity Scale (PSS-78), which is defined as valid for salinity ranges from 2 – 42 PSU.

Do I need to remove batteries before shipping my instrument for a deployment or to Sea-Bird?

Alkaline batteries can be shipped installed in the instrument. See Shipping Batteries for information on shipping instruments with Lithium or Nickel-Metal Hydride (NiMH) batteries.

Do I need to clean the exterior of my instrument before shipping it to Sea-Bird for calibration?

Remove as much biological material and/or anti-foul coatings as possible before shipping. Sea-Bird cannot place an instrument with a large amount of biological material or anti-foul coating on the housing in our calibration bath; if we need to clean the exterior before calibration, we will charge you for this service.

  • To remove barnacles, plug the ends of the conductivity cell to prevent the cleaning solution from getting into the cell. Then soak the entire instrument in white vinegar for a few minutes. After scraping off the barnacles and marine growth, rinse the instrument well with fresh water.
  • To remove anti-foul paint, use a Heavy Duty Scotch-Brite pad (http://www.3m.com/us/home_leisure/scotchbrite/products/scrubbing_scouring.html) or similar scrubbing device.

I want to change the pressure sensor on my CTD, swapping it as needed to get the best data for a given deployment depth. Can I do this myself, or do I need to send the instrument to Sea-Bird?

On most of our instruments, replacement of the pressure sensor should be performed at Sea-Bird. We cannot extend warranty coverage if you replace the pressure sensor yourself.

However, we recognize that you might decide to go ahead and do it yourself because of scheduling/cost issues. Some guidelines follow:

  1. Perform the swap and carefully store the loose sensor on shore in a laboratory or electronics shop environment, not on a ship. The pressure sensor is fairly sensitive to shock, and a loose sensor needs to be stored carefully. Dropping the sensor will break it.
  2. Some soldering and unsoldering is required. Verify that the pressure sensor is mounted properly in your instrument. Properly re-grease and install the o-rings, or the instrument will flood.
  3. Once the sensor is installed, back-fill it with oil. Sea-Bird uses a vacuum-back filling apparatus that makes this job fairly easy. We can provide a drawing showing the general design of the apparatus, which can be modified and constructed by your engineers.
  4. For the most demanding work, calibrate the sensor on a deadweight tester to ensure proper operation and calibration.
  5. Enter the calibration coefficients for the new sensor in:
  • the CTD configuration (.con or .xmlcon) file, using Seasave V7 or SBE Data Processing, and
  • (for an instrument with internally stored calibration coefficients) the CTD EEPROM, using the appropriate terminal program and the appropriate calibration coefficient commands

Note: This discussion does not apply to the SBE 25 (not 25plus), which uses a modular pressure sensor (SBE 29) mounted externally on the CTD. Swap the SBE 29 as desired, use the CC command in Seaterm or SeatermAF to enter the new pressure range and pressure temperature compensation value, and type the calibration coefficients for the new sensor into the CTD configuration (.con or .xmlcon) file in Seasave V7 or SBE Data Processing.

Can I brush-clean and replatinize the conductivity cell myself? How often should this be done?

Brush-cleaning and replatinizing should be performed at Sea-Bird. We cannot extend warranty coverage if you perform this work yourself.

The brush-cleaning and replatinizing process requires specialized equipment and chemicals, and the disassembly of the sensor. If performed incorrectly, you can damage the cell. Additionally, the sensor must be re-calibrated when the work is complete.

Sea-Bird determines whether brush-cleaning and replatinizing is required based upon how far the calibration has drifted from the original calibration. Typically, a conductivity sensor on a profiling CTD requires brush-cleaning and replatinizing every 5 years.

I sent my conductivity sensor to Sea-Bird for calibration, and you also performed a Cleaning and Replatinizing (C &P). You sent the instrument back with 2 sets of calibration data. What does this mean?

The post-cruise calibration contains important information for drift calculations. The post-cruise calibration is performed on the cell as we received it from you, and is an indicator of how much the sensor has drifted in the field. Information from the post-cruise calibration can be used to adjust your data, based on the sensor’s drift over time. See Application Note 31: Computing Temperature and Conductivity Slope and Offset Correction Coefficients from Laboratory Calibrations and Salinity Bottle Samples.

If the sensor has drifted significantly (based on the data from the post-cruise calibration), Sea-Bird performs a C & P to restore the cell to a state similar to the original calibration. After the C & P, the sensor is calibrated again. This calibration serves as the starting point for future data, and for the sensor’s next drift calculation.

The C & P tends to return the cell to its original state. However, there are many subtle factors that may result in the post-C & P calibration not exactly matching the original calibration. Basically, the old platinizing is stripped off and new platinizing is plated on. Anything in this process that alters the cell slightly will result in a difference from the original calibration. We compare the calibration after C & P with the original calibration, not to make any drift analysis, but to make sure we did not drastically alter the cell, or that the cell was not damaged during the C & P process.

How can I tell if the conductivity cell on my CTD is broken?

Conductivity cells are made of glass, which is breakable.

  • If a cell is cracked, it typically causes a salinity shift or erratic data.
  • However, if the crack occurs at the end of the cell, the sensor will continue to function normally until water penetrates the epoxy jacket. Post-cruise calibration results will reveal whether or not water has penetrated the epoxy jacket.

Inspect the cell thoroughly and make sure that it isn’t cracked or abused in any way.

  • (SBE 9plus, 25, or 25plus) If the readings are good at the surface but erratic at depth, it is likely that the problem is in the cable or the connector, not the conductivity cell. Check the connections, making sure that you burp the connectors when you plug them in (see Application Note 57: Connector Care and Cable Installation). Check the cable itself (swap with a spare cable, if available).
  • If the readings are incorrect at the surface but good after a few meters, it is likely that the problem is flow-related. Verify that the pump is working properly. Check the air bleed valve (the white plastic piece in the Y-fitting, which is installed on vertically deployed CTDs) to see if it is clogged; clean out the small hole with a piece of fine wire supplied with your CTD.
  • If the readings are incorrect for the entire cast, there may be an incorrect calibration coefficient or the cell may be cracked.
  • Check the conductivity calibration coefficients in the configuration (.con or .xmlcon) file.
  • Do a frequency check on the conductivity cell. Disconnect the plumbing on the cell. Rinse the cell with distilled or de-ionized water and blow it dry (use your mouth and not compressed air, as there tends to be oil in the air lines on ships). With the cell completely dry, check the frequency reading. It should read within a few tenths of a Hz of the 0 reading on your Calibration Sheet. If it does not, something is wrong with the cell and it needs to be repaired.

What are the major steps involved in deploying a moored instrument?

Application Note 83: Deployment of Moored Instruments contains a checklist, which is intended as a guideline to assist you in developing a checklist specific to your operation and instrument setup.

What are the recommended practices for connectors - mating and unmating, cleaning corrosion, and replacing?

Mating and Unmating Connectors:

It is important to prepare and mate connectors correctly, both in terms of the costs to repair them and to preserve data quality. Leaking connectors cause noisy data and even potential system shutdowns. Application Note 57: Connector Care and Cable Installation describes the proper care and installation of connectors for Sea-Bird instruments. The Application Note covers connector cleaning and cable or dummy plug installation, locking sleeve installation, and cold weather tips.

Checking for Leakage and Cleaning Corrosion on Connectors:

If there has been leakage, it will show up as green-colored corrosion product. Performing the following steps can usually reverse the effect of the leak:

  1. Thoroughly clean the connector with water, followed by alcohol.
  2. Give the connector surfaces a light coating of silicon grease.

Re-mate the connectors properly — see Application Note 57: Connector Care and Cable Installation and 9-minute video covering O-ring, connector, and cable maintenance.

Replacing Connectors:

  • The main concern when replacing a bulkhead connector is that the o-rings on the connector and end cap must be prepared and installed correctly; if they are not, the instrument will flood. See the question below for general procedure on handling o-rings.
  • Use a thread-locking compound on the connector threads to prevent the new connector from loosening, which could also lead to flooding.
  • If the cell guard must be removed to open the instrument, take extra care not to break the glass conductivity cell.

What is an Anti-Foulant Device? Does it affect the conductivity cell calibration? How often should I replace it? Does it require special handling?

The Anti-Foulant Device is an expendable device that is installed on each end of the conductivity cell, so that any water that enters the cell is treated. Anti-Foulant Devices are typically used with moored instruments (SBE 16, 16plus, 16plus-IM, 16plus V2, 16plus-IM V2, 37-SM, 37-SMP, 37-SMP-IDO, 37-SMP-ODO, 37-SI, 37-SIP, 37-SIP-IDO, 37-IM, 37-IMP, 37-IMP-IDO, 37-IMP-ODO), thermosalinographs (SBE 21 and 45), glider CTDs (Glider Payload CTD), moored profilers (SBE 52-MP), and drifters (SBE 41/41CP Argo float CTDs), and optionally with SBE 19plus, 19plus V2, and 49 profilers.

Anti-Foulant Devices have no effect on the calibration, because they do not affect the geometry of the conductivity cell in any way. The Anti-Foulant Devices are mounted at either end of the conductivity cell. For an in-depth explanation of how Sea-Bird makes the conductivity measurement, see Conductivity Sensors for Moored and Autonomous Operation.

Useful deployment life varies, depending on several factors. We recommend that customers consider more frequent anti-foulant replacement when high biological activity and strong current flow (greater dilution of the anti-foulant concentration) are present. Moored instruments in high growth and strong dilution environments have been known to obtain a few months of quality data, while drifters that operate in non-photic, less turbid deep ocean environments may achieve years of quality data. Experience may be the strongest determining factor in specific deployment environments. Sea-Bird recommends that you keep track of how long the devices have been deployed, to allow you to purchase and replace the devices when needed.

Note that the anti-foulant device does not actually dissolve, so there is no way to visually determine if the anti-foulant device is still effective.

The cost of the anti-foulant devices is small compared to the deployment costs, so we recommend that you replace the devices before each deployment. This will provide the maximum bio-fouling protection, resulting in long-term data quality. 

Shelf Life and Storage: The best way to store Anti-Foulant Devices is in an air-tight, opaque container. The rate of release of anti-foulant is based on saturation of the environment. The anti-foulant will release until the environment is fully saturated (100% saturated) and then it will no longer release any anti-foulant. So if you keep Anti-Foulant Devices sealed well in an air-tight container, theoretically they will stay good for extended periods of time. Exposure to direct sunlight can also affect the release of anti-foulant; we recommend storage in an opaque container.

Handling:

  • For details, refer to the Material Safety Data Sheet, enclosed with the shipment and available on our MSDS page.
  • Anti-Foulant Devices are not classified by the U.S. DOT or the IATA as hazardous material.

What are the recommended practices for storing sensors at low temperatures, and deploying at low temperatures or in frazil or pancake ice?

General

Large numbers of Sea-Bird conductivity instruments have been used in Arctic and Antarctic programs.

Special accommodation to keep temperature, conductivity, oxygen, and optical sensors at or above 0 C is advised. Often, the CTD is brought inside protective doors between casts to achieve this.

Conductivity Cell

When freezing is possible, we recommend that the conductivity sensor be stored dry. Remove larger droplets of water by blowing through the cell. Do not use compressed air, which typically contains oil vapor. Attach a length of Tygon tubing to each end of the conductivity cell to close the cell ends. See Application Note 2D: Instructions for Care and Cleaning of Conductivity Cells for details.

There are several considerations to weigh when contemplating deployments at low temperatures in general, and in frazil or pancake ice:

  • Ensure that the instrument is at or above water temperature before it is deployed. If the cell gets colder than 0 to -2 ºC while on deck, when it enters the water a layer of ice forms inside the cell as the cell warms to ocean temperature. If ice forms inside the conductivity cell, measurements will be low of correct until the ice layer melts and disappears. Thin layers of ice will not hurt the conductivity cell, but repeated ice formation on the electrodes will degrade the conductivity calibration (at levels of 0.001 to 0.020 psu) and thicker layers of ice can lead to glass fracture and permanent damage of the cell.
  • For accurate measurements, keep ice out of the sensing region of the conductivity cell. The conductivity measurement involves determining the electrical resistance of the water inside the sensor. Ice is essentially a non-conductor. To the extent that ice displaces the water, the conductivity will register (very) misleadingly low. Some type of screening is necessary to keep ice out of the cell. This is relatively easy to arrange for the Sea-Bird conductivity cell, which is an electrode-type cell, because its sensing region is totally inside a long tube; plastic mesh could be positioned at each end and would have zero effect on accuracy and stability.

The above considerations apply to all known conductivity sensor types, whether electrode or inductive types. 

If deploying at low temperatures but no surface frazil or pancake ice is present, rinse the conductivity cell in one of the following salty solutions (salty water depresses the freezing point) to prevent freezing during deployment. But this does not mean you can store the cell in one of these solutions outside . . . it will freeze.

  • Solution of 1% Triton in sterile seawater (use 0.5-micron filtered seawater or boiled seawater),   or
  • Brine solution (distilled seawater or homemade salt solution that is higher than 35 psu in salinity).

Note that there is still a risk of forming ice inside the conductivity cell if deploying through frazil or pancake ice on the surface, if the freezing point of the salt water is the same as the water temperature. Therefore, we recommend that you deploy the conductivity cell in a dry state for these deployments.

Commercially available alcohol or glycol antifreezes contain trace amounts of oils that will coat the conductivity cell and the electrodes, causing a calibration shift, and consequently result in errors in the data. Do not use alcohol or glycol in the conductivity cell.

Temperature Sensor

In general, neither the accuracy of the temperature measurement nor the survival of the temperature sensor will be affected by ice.

Oxygen Sensor

For the SBE 43 and SBE 63 Dissolved Oxygen sensor, avoid prolonged exposure to freezing temperature, including during shipment. Do not store the with water (fresh or seawater), Triton solution, alcohol, or glycol in the plenum. The best precaution is to keep the sensor indoors or in some shelter out of the cold weather.

Does it matter if I deploy my moored instrument, which includes a conductivity sensor, in a horizontal or vertical position?

Yes, vertical is usually preferable. In the presence of consistent currents and suspended sediment, we have seen instances where a horizontal conductivity cell is scoured by the abrasive effect of the flow. When scouring is particularly intense, the electrodes can be stripped of their electroplated platinum-black coating, driving the calibration toward fresher readings. Sedimentation (silting) in the cell also drives the readings fresh of correct.

Mounting the instrument vertically avoids abrasive flow and sediment build-up while allowing wave motions and Bernoulli pressures to flush the cell.

Note that some moored sensors (SBE 37-SIP37-SIP-IDO, 37-SMP37-SMP-IDO37-SMP-ODO37-IMP37-IMP-IDO37-IMP-ODO) have a recommended orientation because of their u-shaped plumbing configuration. Refer to the instrument manual for details.

Is it necessary to put my instrument in water to test it? Will I destroy the conductivity cell if I test it in air?

It is not necessary to put the instrument in water to test it. It will not hurt the conductivity cell to be in air.

If there is a pump on the instrument, it should not be run for extended periods in air.

  • Profiling instruments (SBE 9plus, 19, 19plus, 19plus V2, 25, 25plus, 49) and some moored instruments (all pumped MicroCATs with integral dissolved oxygen (DO), and pumped MicroCATs without DO with firmware 3.0 and later) do not turn on the pump unless the conductivity frequency is above a specified minimum value (minimum value is hard-wired in 9plus, user-programmable in other instruments). This prevents the pump from turning on in air. See the instrument manual for details.
  • If your instrument does not check for conductivity frequency before turning on the pump: 
    - For moored SeaCATs (16, 16plus, 16plus-IM, 16plus V2, 16plus-IM V2): Disconnect the pump cable for the test. 
    - For older pumped MicroCATs: orient the MicroCAT to provide an upright U-shape for the plumbing. Then fill the inside of the pump head with water via the pump exhaust tubing; this will provide enough lubrication to prevent pump damage during brief testing.

How should I handle my CTD to avoid cracking the conductivity cell?

Shipping: Sea-Bird carefully packs the CTD in foam for shipping. If you are shipping the CTD or conductivity sensor, carefully pack the instrument using the original crate and packing materials, or suitable substitutes.

Use: Cracks at the C-Duct end of the conductivity cell are most often caused by:

  • Hitting the bottom, which can cause the T-C Duct to flex, resulting in cracking at the end of the cell.
  • Removing the soaker tube from the T-C duct in a rough manner, which also causes the T-C Duct to flex. Pulling the soaker tube off at an angle can be especially damaging over time to the cell. Pull the soaker tube off straight down and gently.
  • Improper disassembly of the T-C ducted temperature and conductivity sensors (SBE 25, 25plus, and 9plus) when removing them for shipment to Sea-Bird for calibration. See Shipping SBE 9plus, 25, and 25plus Temperature and Conductivity Sensors for the correct procedure.

Note: If a Tygon tube attached to the conductivity cell has dried out, yellowed, or become difficult to remove, slice (with a razor knife or blade) and peel the tube off of the conductivity cell rather than twisting or pulling the tube off.

Can I use a pressure sensor above its rated pressure?

Digiquartz pressure sensors are used in the SBE 9plus, 53, and 54. The SBE 16plus V2, 16plus-IM V2, 19plus V2, and 26plus can be equipped with either a Druck pressure sensor or a Digiquartz pressure sensor. All other instruments that include pressure use a Druck pressure sensor.

  • The overpressure rating for a Digiquartz (as stated by Paroscientific) is 1.2 * full scale. The sensor will provide data values above 100% of rated full scale; however, Sea-Bird does not calibrate beyond the rated full scale.
  • The overpressure rating for a Druck (as stated by Druck) is 1.5 * full scale. The sensor will provide data values above 100% of rated full scale; however, Sea-Bird does not calibrate beyond the rated full scale.

Note: If you use the instrument above the rated range, you do so at your own risk; the product will not be covered under warranty.

What are the typical data processing steps recommended for each instrument?

Section 3: Typical Data Processing Sequences in the SBE Data Processing manual provides typical data processing sequences for our profiling CTDs, moored CTDs, and thermosalinographs. Typical values for aligning, filtering, etc. are provided in the sections detailing each module of the software. This information is also documented in the software's Help file. To download the software and/or manual, go to SBE Data Processing.

How do instruments handle external power if internal batteries are installed?

Most Sea-Bird instruments that are designed to be powered internally or externally incorporate diode or'd circuitry, allowing only the voltage that has the greater potential to power the instrument. You can power the instrument externally without running down the internal batteries. This allows you to lab test using external power that has higher voltage than the internal batteries, and then deploy using internal power, knowing that the internal batteries are fresh.

For the SBE 25plus, if external power of 14 volts or higher is applied, the 25plus runs off of the external power, even if the main battery voltage is higher.

How should I pick the pressure sensor range for my CTD? Would the highest range give me the most flexibility in using the CTD?

While the highest range does give you the most flexibility in using the CTD, it is at the expense of accuracy and resolution. It is advantageous to use the lowest range pressure sensor compatible with your intended maximum operating depth, because accuracy and resolution are proportional to the pressure sensor's full scale range. For example, the SBE 9plus pressure sensor has initial accuracy of 0.015% of full scale, and resolution of 0.001% of full scale. Comparing a 2000 psia (1400 meter) and 6000 psia (4200 meter) pressure sensor:

  • 1400 meter pressure sensor ‑ initial accuracy is 0.21 meters and resolution is 0.014 meters
  • 4200 meter pressure sensor ‑ initial accuracy is 0.63 meters and resolution is 0.042 meters

What is the maximum cable length for real-time RS-232 data?

Cable length is one of the most misunderstood items in the RS-232 world. The RS-232 standard was originally developed decades ago for a 19200 baud rate, and defines the maximum cable length as 50 feet, or the cable length equal to a capacitance of 2500 pF. The capacitance rule is often forgotten; using a cable with low capacitance allows you to span longer distances without going beyond the limitations of the standard. Also, the maximum cable length mentioned in the standard is based on 19200 baud rate; if baud is reduced by a factor of 2 or 4, the maximum length increases dramatically. Using typical underwater cables, allowable combinations of cable length and baud rate for Sea-Bird instruments communicating with RS-232 are shown below:

Maximum Cable Length (meters) Maximum Baud Rate*
1600 600
800 1200
400 2400
200 4800
100 9600
50 19,200
25 38,400
16 57,600
8 115,200

*Note: Consult instrument manual for baud rates supported for your instrument.

 

Family Model . Pressure Sensor/Range
SEAPHOX   . 0 – none
      1 – 20 m strain gauge
      2 – 100 m strain gauge

Example: SEAPHOX.1 is a SeapHOx with 20 m strain gauge pressure sensor. See table below for description of each selection:

PART # DESCRIPTION NOTES
SEAPHOX

 

 
SeaPHOx Configuration Selections MUST SELECT ONE
SEAPHOX.0 SeapHOx Ocean CT (no pressure sensor) + pH + DO sensor. Integrates the Satlantic SeaFET (v 1.0) Ocean pH sensor with flow cell and the Sea-Bird SBE 37SMP-ODO MicroCAT C and T Recorder with Integrated SBE 63 Optical Dissolved Oxygen sensor (ODO) with titanium housing. 50 meter depth rating. All sensors using wet-pluggable (MCBH) connector. Includes SeapHOx integration kit (mount brackets, flow tube adapter, Y-cable, USB programming cable, hardware). Pressure sensor is installed in MicroCAT end cap (behind mount clamp), & is not field replaceable / swappable. While highest pressure rating gives you most flexibility in using MicroCAT, it is at expense of accuracy & resolution. It is advantageous to use lowest range pressure sensor compatible with your intended maximum operating depth, because accuracy & resolution are proportional to pressure sensor's full scale range. For example, comparing 20 & 100 m sensors:
  • 20 m sensor:
    initial accuracy = 0.02 m (= 0.1% * 20 m),
    resolution = 0.0004 m (= 0.002% * 20 m)
  • 100 m sensor:
    initial accuracy = 0.1 m (= 0.1% * 100 m),
    resolution = 0.002 m (= 0.002% * 100 m)
SEAPHOX.1 SeapHOx Ocean CTD + pH + DO sensor. Integrates the Satlantic SeaFET (v 1.0) Ocean pH sensor with flow cell and the Sea-Bird SBE 37SMP-ODO MicroCAT CTD with 20 meter pressure sensor, with Integrated SBE 63 Optical Dissolved Oxygen sensor (ODO) with titanium housing. 20 meter depth rating. All sensors using wet-pluggable (MCBH) connector. Includes SeapHOx integration kit (mount brackets, flow tube adapter, Y-cable, USB programming cable, hardware).
SEAPHOX.2 SeapHOx Ocean CTD + pH + DO sensor. Integrates the Satlantic SeaFET (v 1.0) Ocean pH sensor with flow cell and the Sea-Bird SBE 37SMP-ODO MicroCAT CTD with 100 meter pressure sensor, with Integrated SBE 63 Optical Dissolved Oxygen sensor (ODO) with titanium housing. 50 meter depth rating. All sensors using wet-pluggable (MCBH) connector.  Includes SeapHOx integration kit (mount brackets, flow tube adapter, Y-cable, USB programming cable, hardware).
SBE 37-SMP Mooring Clamp Wire Size Selections (Specify clamp to match O.D. of mooring wire jacket; cannot accommodate wire size greater than 1.2 in. diameter) — MUST SELECT ONE
37SMP-1a Wire guide & mounting clamp for 1/4 in. diameter mooring wire Cable fits loosely through wire guide, & is clamped only at mounting clamp. See document 67219.

Thread for clamping to mooring cable:

  • 37SMP-1a (1/4 in. diameter): 1/4-28 UNF
  • 37SMP-1b (5/16 in. diameter): 5/16-24 UNF
  • 37SMP-1c (3/8 in. diameter): 3/8-24 UNF
  • 37SMP-1d (1/2 in. diameter): 9/16-12 UNC
  • 37SMP-1e (6 mm diameter): 1/4-20 UNF
  • 37SMP-1f (8 mm diameter): 5/16-24 UNF
  • 37SMP-1g (10 mm diameter): 7/16 -14 UNF
  • 37SMP-1h (12 mm diameter): 1/2-13 UNF

Clamp Size Note: Mooring wire is typically specified by wire size, not by outer diameter (O.D.) of the mooring wire jacket. Verify the wire jacket O.D. before selecting the clamp size. The clamp size must be less than or equal to the wire jacket O.D. but larger than the wire diameter. For example, Mooring System Inc.’s specifications for 3x19 wire rope (in 2016) are as follows:

Wire Diameter Jacket Diameter Recommended Sea-Bird Clamp
3/16 inch (5.0 mm) 0.255 inch (6.5 mm) 1/4 inch
1/4 inch (6.5 mm) 0.330 inch (8.4 mm) 5/16 inch
5/16 inch (8.0 mm) 0.392 inch (9.9 mm) 3/8 inch
3/8 inch (9.5 mm) 0.453 inch (11.5 mm) 10 mm (0.394 inch)
7/16 inch (11.1 mm) 0.5 inch (12.7 mm) 1/2 inch

For mounting on a rope, verify the rope outer diameter, and select a clamp smaller than the rope O.D. to account for the rope compressibility (for example, for 5/16 inch rope, select a ¼ inch clamp; a 5/16 inch clamp will be too large).

37SMP-1b Wire guide & mounting clamp for 5/16 in. diameter mooring wire
37SMP-1c Wire guide & mounting clamp for 3/8 in. diameter mooring wire
37SMP-1d Wire guide & mounting clamp for 1/2 in. diameter mooring wire
37SMP-1e Wire guide & mounting clamp for 6 mm diameter mooring wire
37SMP-1f Wire guide & mounting clamp for 8 mm diameter mooring wire
37SMP-1g Wire guide & mounting clamp for 10 mm diameter mooring wire
37SMP-1h Wire guide & mounting clamp for 12 mm diameter mooring wire
37SMP-1j Alternative mounting - plastic brackets to provide through-bolt mounting to a flat surface See document 67218.
SeapHOx Spares & Accessories - See selections in SBE 37-SMP and SeaFET
SeapHOx Integration Kits
50618 SeapHOx integration kit (mount brackets, flow tube adapter, Y-cable (MCBH SBE 37 connector), fasteners These kits are intended for use by customers who already own a SeaFET and/or a MicroCAT and want to perform the integration without sending the instrument(s) back to the factory.
50637 SeapHOx integration kit (mount brackets, flow tube adapter, Y-cable (XSG SBE 37 connector), fasteners

From Satlantic (SeaFET)

Flow Cell

It is essential that the SeaFET™ sensing elements remain immersed in seawater at all times.  For this reason, the SeaFET™ is equipped with a a resealable flow cell. The flow cell can be fitted with brass penetrating fittings and copper-nickel tubes to which one can attach flexible tubing or hoses for flow-through applications (Image shows one fitting attached and one not).

From Sea-Bird Electronics (SBE 37-SMP-ODO MicroCAT CTD+DO)

Many cables, mount kits, and spare parts can be ordered online.

Cables

RS-232 or RS-485 versions

  • 801385 To computer COM port with power leads (from XSG connector), 2.4 m, DN 32277 (0)
  • 801376 To computer COM port with 9V connector (from XSG connector), 2.4 m, DN 32604 (0)
  • 801206 To computer COM port with power leads (from Wet-pluggable connector), 2.4 m, DN 32366 (0)
  • 801263 To computer COM port with 9V connector (from Wet-pluggable connector), 2.4 m, DN 32490 (0)

Dual SDI-12 / RS-232 version

  • 802220 To computer COM port with power leads and leads to SDI-12 (from Wet-pluggable connector), 2.4 m, DN 33733 (0)

Mount Kits

Mounted to Flat Surface

  • 50479 Thru Bolt Mounting Clamp Kit (document 67218 (0))

Mounted to Mooring Cable (document 67219 (0) for all sizes)

  • 50465 Cable Clamp Kit, 1/4-inch diameter 
  • 50469 Cable Clamp Kit, 5/16-inch diameter
  • 50470 Cable Clamp Kit, 3/8-inch diameter
  • 50473 Cable Clamp Kit, 1/2-inch diameter
  • 50467 Cable Clamp Kit, 6-mm diameter
  • 50468 Cable Clamp Kit, 7-mm diameter
  • 50471 Cable Clamp Kit, 10-mm diameter
  • 50472 Cable Clamp Kit, 11-mm diameter
  • 50474 Cable Clamp Kit, 12-mm diameter

Clamp Size Note: Mooring wire is typically specified by wire size, not by outer diameter (O.D.) of the mooring wire jacket. Verify the wire jacket O.D. before selecting the clamp size. The clamp size must be less than or equal to the wire jacket O.D. but larger than the wire diameter. For example, Mooring System Inc.’s specifications for 3x19 wire rope (in 2016) are as follows:

Wire Diameter Jacket Diameter Recommended Sea-Bird Clamp
3/16 inch (5.0 mm) 0.255 inch (6.5 mm) 1/4 inch
1/4 inch (6.5 mm) 0.330 inch (8.4 mm) 5/16 inch
5/16 inch (8.0 mm) 0.392 inch (9.9 mm) 3/8 inch
3/8 inch (9.5 mm) 0.453 inch (11.5 mm) 10 mm (0.394 inch)
7/16 inch (11.1 mm) 0.5 inch (12.7 mm) 1/2 inch

For mounting on a rope, verify the rope outer diameter, and select a clamp smaller than the rope O.D. to account for the rope compressibility (for example, for 5/16 inch rope, select a ¼ inch clamp; a 5/16 inch clamp will be too large).

Spare Parts

Batteries

  • 801863 Yellow battery holder (14V nominal, Version 2) for SBE 37 (SM, SMP, IM, IMP with firmware version > 4.0) and SBE 37 with oxygen (SMP-IDO, IMP-IDO, SMP-ODO, IMP-ODO), for use with twelve 3.6V AA lithium cells
  • 50441 SBE 37 and 44 lithium batteries, package of twelve 3.6V AA cells (Saft LS 14500)

Hardware & O-Ring Kits

  • 60056 Hardware & O-ring kit (document 67212 (0))

Miscellaneous

  • 801542 AF24173 Anti-Foulant Device (pair, bagged, labeled for shipping)

Compare  Moored / Time Series Recording Instruments

SBE Measures
(C, T, P)
Auxiliary Sensors Memory Power Communication Real-Time
Data
Comments
Internal External
SBE 16plus V2 SeaCAT C-T (P) Recorder C, T, P* 6 A/D; 1 RS-232 64 Mb RS-232 Optional pump
SBE 16plus SeaCAT C-T (P) Recorder
C, T, P* 4 A/D; optional RS-232 or PAR 8 Mb RS-232 or -485 Replaced by SBE 16plus V2 in 2008
SBE 16 SeaCAT C-T (P) Recorder
C, T, P* 4 A/D 1 Mb RS-232 Replaced by SBE 16plus in 2001
SBE 16plus-IM V2 SeaCAT C-T (P) Recorder C, T, P* 6 A/D; 1 RS-232 64 Mb   Inductive Modem Optional pump
SBE 16plus-IM SeaCAT C-T (P) Recorder
C, T, P* 4 A/D; optional RS-232 or PAR 8 Mb   Inductive Modem Replaced by SBE 16plus-IM V2 in 2008
SBE 19plus V2 SeaCAT Profiler CTD C, T, P 6 A/D; 
1 RS-232
64 Mb RS-232 Programmable mode — profiling or moored
SBE 19plus SeaCAT Profiler CTD
C, T, P 4 A/D; optional PAR 8 Mb RS-232 Replaced by SBE 19plus V2 in 2008
SBE 19 SeaCAT Profiler CTD
C, T, P 4 A/D 1 - 8 Mb RS-232 Replaced by SBE 19plus in 2001
SBE 37-SM MicroCAT C-T (P) Recorder C, T, P*   8 Mb RS-232 or -485  
SBE 37-SMP MicroCAT C-T (P) Recorder C, T, P*   8 Mb RS-232, RS-485, or SDI-12 Integral pump
SBE 37-SMP-IDO MicroCAT C-T-DO (P) Recorder C, T, P* Integrated DO 8 Mb RS-232 or -485 Integral pump; Replaced by SBE 37-SMP-ODO in 2014
SBE 37-SMP-ODO MicroCAT C-T-DO (P) Recorder C, T, P* Integrated Optical DO 8 Mb RS-232, RS-485, or SDI-12 Integral pump
SBE 37-IM MicroCAT C-T (P) Recorder C, T, P*   8 Mb   Inductive modem  
SBE 37-IMP MicroCAT C-T (P) Recorder C, T, P*   8 Mb   Inductive modem Integral pump
SBE 37-IMP-IDO MicroCAT C-T-DO (P) Recorder C, T, P* Integrated DO 8 Mb   Inductive modem Integral pump; Replaced by SBE 37-IMP-ODO in 2014
SBE 37-IMP-ODO MicroCAT C-T-DO (P) Recorder C, T, P* Integrated Optical DO 8 Mb   Inductive modem Integral pump
SBE 37-SI MicroCAT C-T (P) Recorder C, T, P*   8 Mb   RS-232 or -485  
SBE 37-SIP MicroCAT C-T (P) Recorder C, T, P*   8 Mb   RS-232 or -485 Integral pump

SBE 37-SIP-IDO MicroCAT C-T-DO (P) Sensor

C, T, P* Integrated DO 8 Mb   RS-232 or -485 Integral pump
SBE 39plus Temperature (P) Recorder T, P*   64 Mb Optional USB & RS-232 Optional  
SBE 39 Temperature (P) Recorder
T, P*   32 Mb Optional RS-232 Optional Replaced by SBE 39plus in 2014
SBE 39plus-IM Temperature (P) Recorder T, P*   64 Mb   Inductive Modem & USB  
SBE 39-IM Temperature (P) Recorder T, P*   32 Mb   Inductive modem Replaced by SBE 39plus-IM in 2016
SBE 56 Temperature Logger T   64 Mb   USB    
SBE 26plus Seagauge Wave & Tide Recorder T, P C optional 32 Mb RS-232
(tides, waves, & wave statistics)
Wave & tide recorder
SBE 26 Seagauge Wave & Tide Recorder
T, P C optional 8 Mb RS-232   Replaced by SBE 26plus in 2004
SBE 53 BPR Bottom Pressure Recorder T, P C optional 32 Mb RS-232 Bottom pressure recorder
SBE 54 Tsunameter Tsunami Pressure Sensor T, P   128 Mb Optional RS-232 Tsunami pressure sensor

C = conductivity, T = temperature, P = pressure, DO = dissolved oxygen
* = optional
products are no longer in production. Follow the links above to the product page to retrieve manuals and application notes for these older products.