Seabird

SBE 37 SMP-ODO with Plastic Housing, No Pressure Sensor, MCBH Connector, RS-232, 600 meter ODO Sensor, Includes Accessories

SBE 37 SMP-ODO with Plastic Housing, No Pressure Sensor, MCBH Connector, RS-232, 600 meter ODO Sensor, Includes Accessories
Product #: 37SMP.10202S
Unit Price Get Quote

Optimal Moored CTD

Moored Conductivity, Temperature, and Pressure (optional), at user-programmable 6-sec to 6-hour intervals.

Flexible Deployment Options

Internal memory and battery pack (can be powered externally).

Integral Pump

Pump runs for 1 second for each sample, providing improved conductivity response and bio-fouling protection.

Optical Dissolved Oxygen Sensor

Integrated SBE 63 Optical Dissolved Oxygen Sensor. Adaptive Pump Control for high-accuracy oxygen data.

Specifications

Communication: RS-232
Conductivity Accuracy: ± 0.0003 S/m (0.003 mS/cm)
Conductivity Measurement Range: 0 - 7 S/m (0 - 70 mS/cm)
Conductivity resolution: 0.00001 S/m (0.0001 mS/cm)
Conductivity Typical Stability: 0.0003 S/m (0.003 mS/cm) per month
Connector: MCBH
Dissolved Oxygen Accuracy: larger of ± 3 µmol/kg (equivalent to 0.07 ml/L or 0.1 mg/L) or ± 2%
Dissolved Oxygen Range: 120% of surface saturation in all natural waters (fresh and salt)
Dissolved Oxygen Resolution: 0.2 µmol/kg
Dissolved Oxygen Typical Stability: Sample-based drift < 1 µmol/kg/100,000 samples (20 °C)
Housing Material: Plastic
Internal Batteries: 10.6 Amp-hour (nominal) battery pack (derated for calculations)
Power input: 0.25 Amps at 9 - 24 VDC
Pressure Initial Accuracy: N/A
Pressure Resolution: N/A
Pressure Sensor/Range: No Pressure Sensor
PressureTypical Stability: N/A
Pumps: Yes
Sensors: 600 meter ODO Sensor
Temperature Accuracy: ± 0.002 °C (-5 to +35 °C); ± 0.01 (+35 to +45 °C)
Temperature Range: -5 to +45 °C
Temperature Resolution: 0.0001 °C
Temperature Stability: 0.0002 °C per month
Weight: 3.4 kg in air, 1.5 kg in water.